Paper ID S003 : Effect of keeper after MRI irradiation on attractive force


Posted Messages (Last post: Fri Mar 13 15:09:40 2020)
[0001] [0002] [0003] 
[Go to Form]

[0001]
Yukyo Takada (Tohoku University Graduate School of Dentistry )

I evaluate this presentation clinically as an important and practical study.

Please tell us concretely why the large variation occurred in the attractive forces of magnetic assemblies?

[0002]
Shin-ichi Masumi (Kyushu Dental University )

The cause is considered to be the following two technical errors.
First, when the keeper is bonded to the duplicated lower non-magnetic table, a slight horizontal shift occurs between the keeper and the magnetic assembly.
Another possibility is that a slight displacement occurred when fixing the duplicated lower non-magnetic table with the keeper bonded to the universal tensile tester.

[0003]
Yukyo Takada (Tohoku University Graduate School of Dentistry )

Thank you for your response. Although I understand very difficult measurements, do you find a good idea or way to reduce these technical errors after finished these measurements?